Juergensen’s Threat Matrix
Diver forgets to turn on unit:
¾ Answer: Wet Switches
Wet Switches Fail:
¾ Answer: Pressure Transducer will activate unit at 1m pressure
Diver sets unit to Open Circuit, but is still breathing the loop.
¾ Answer: Solenoid Override at 0.19 PO
2
Diver sets unit to Manual Control, but forgets to add O2:
¾ Answer: Solenoid Override at 0.19 PO
2
Diver ignores Primary and Secondary Display:
¾ Answer: Add HUD/DIVA
Diver ignores or is unaware of DIVA LED Red Warning of PO
2
Danger:
¾ Answer: Trigger Vibrator at 1.8 and above, or 0.19 and below.
Diver ignores Vibrator and LED:
¾ Answer:
Primary Red or Green LED firing
• Red for Low ppO2
• Green for High ppO2
Secondary Red LED firing
Diver ignores LED's in Primary and Secondary:
Answer: Backlights in BOTH handsets begin to flash. This is
highly visible to both the diver, and any nearby divers.