O2ptima Manual –User Guide – Rev 1.2 Page 18 of 26
__________________________________________________________________
Proprietary Information - Not to be released without written authorization
Asseer & Associates, 5609 Power Road, Ottawa, Ontario, K1G 3N4
IMPORTANT!!! When re-assembling the DSV, extreme care should be taken to ensure
the valves have been put in the proper orientation for air loop flow. The long hose with red
connector should be on the inhalation side and the long hose with black connector should
be on the exhalation side. A mushroom valve check should perform after reassembly.
3.5. O-Rings
The O2ptima CCR has multiple o-rings contained within the breathing loop. All o-rings
should be changed on a yearly basis or sooner if degradation is seen. The canister has nine
(9) o-rings and the breathing loop has sixteen (16) o-rings that need to be periodically
inspected and replaced each year. The reason for such a high o-ring count in the O2ptima
CCR is that the majority of fittings have double o-ring fittings.
3.6. Batteries
The recommended battery for the O2ptima CCR electronics is the 3.6 Volt Saft Lithium
Ion (Saft P/N: LS14500-10). Many factors affect battery life such as the frequency of the
solenoid firing and back lighting for the electronics. It is recommended that the battery be
replaced every 6 hours of diving but careful attention should be used if dive profiles require
a high frequency of solenoid firing or long duration of back lighting.
If the unit is not going to be used for an extended period of time (more than 24 hours), it is
recommended to remove the batteries to conserve power.
WARNING! WHEN THE BATTERIES ARE REMOVED FROM THE HANDSETS,
ALL TISSUE INFORMATION IS LOST. A SURFACE INTERVAL OF 24 HRS IS
REQUIRED PRIOR TO USING THE HAMMERHEAD ELECTRONICS TO PROVIDE
DECOMPRESSION INFORMATION.
Additional information on batteries can be found in the HammerHead manual.
3.7. Wiring Harness
The wiring harness relays the electrical signal from each of the cells to the main connection
block inside the head. It is critical that a good electrical contact is made between the
connection block and the head. Improper connection can lead to erroneous oxygen levels
or the inability to calibrate a specific cell.
The banana jack connectors and receptacles should be cleaned every 20 hours of use or any
time excessive moisture or flooding has occurred. The wiring harness should be replaced